Ground-Based Microwave Radiometer Calibration: an Overview
Keywords:
Cryogenic calibration, Atmospheric profiling, Microwave radiometers, Ground-based radiometryAbstract
This paper intends to briefly present some basic concepts on the microwave radiometry and radiometer calibration research in remote sensing applications and demonstrate results and analysis of the cryogenic calibration of a microwave ground-based radiometer currently deployed in scientific campaigns in Brazil. The equipment described in this text operates at 22 – 30 GHz and at 51 – 59 GHz frequency ranges and uses as the calibration standard a target cooled by liquid nitrogen. Since an accurate calibration (with observation errors below 0.5 K) is important to provide confidence in the retrieval of vertical temperature and humidity profiles, this work aims also to comment on some effects of the errors in calibration procedure on the atmospheric parameters of interest.Downloads
Published
2018-07-18
Issue
Section
Review Articles
License
This work is licensed under a Creative Commons — Attribution 4.0 International — CC BY 4.0. Authors are free to Share (copy and redistribute the material in any medium or format) and Adapt (remix, transform, and build upon the material for any purpose, even commercially). JATM allow the authors to retain publishing rights without restrictions.